Abstract

3D printing technology revolutionized the development of electrochemical sensors allowing the fabrication of custom devices that can be adapted to meet specific applications. This study describes the development of a miniaturized electrochemical cell containing integrated electrodes manufactured entirely by 3D printing for the sequential detection of steroid hormones. The electrochemical sensor was manufactured in a single step by combining two types of thermoplastic filaments. Polylactic acid-Carbon Black (PLA-CB) was used in the electrode region and acrylonitrile butadiene styrene (ABS) was used in the insulating region. Chemical/electrochemical treatments were used to remove the inert polymer matrix and expose the electroactive sites of the CB electrodes. These pre-treatment procedures significantly enhanced the electrochemical response of the printed sensor. The analytical performance of the activated 3D printed electrochemical cell was demonstrated in the sequential detection of estradiol (E2) and progesterone (P4) in artificial urine. The electrochemical cell offered good repeatability (RSD = 3.1%) and reproducibility (RSD = 10.7%) between analyses reaching detection limits of 0.11 and 17.8 µmol L-1 for E2 and P4, respectively. In this sense, these results demonstrate that the use of 3D printing paves the way for the fast and economical fabrication of miniaturized electrochemical cells for hormone detection in biological fluids.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call