Abstract
The study investigates the short-term behaviour of the acetabular construct following revision hip arthroplasty, carried out using the Slooff–Ling impaction grafting technique; using 3D finite element analyses. An elasto-plastic material model is used to describe the constitutive behaviour of morsellised cortico-cancellous bone (MCB) graft, since it has been shown that MCB undergoes significant plastic deformation under normal physiological loads. Based on previous experimental studies carried out by the authors and others, MCB is modelled using non-linear elasticity and Drucker Prager Cap (DPC) plasticity. Loading associated with walking, sitting down, and standing up is applied to the acetabular cup through a femoral head using smooth sliding surfaces. The analyses yield distinctive patterns of migration and rotation due to different activities. These are found to be similar to those observed in the clinical setting.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Computer Methods in Biomechanics and Biomedical Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.