Although a variety of timing techniques and data acquisition strategies have been used for three-dimensional contrast-enhanced MR angiography, many are still limited by inadequate overall reliability, limited spatial resolution, or complexity. A technique is presented in this work in which contrast arrival is detected in the targetted vasculature in real time using MR fluoroscopy. Upon detection the operator triggers a 3D MR angiographic acquisition which uses an elliptical centric view order. It is shown that the view order intrinsically provides a high degree of venous suppression which in turn allows acquisition times of 30 seconds or longer, permitting high spatial resolution. The reliability of fluoroscopic triggering in bolus detection is shown to be over 90%. The technique provides high quality contrast-enhanced MR angiograms for many vascular regions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.