Abstract
To develop and validate a three-dimensional (3D) single breath-hold, projection reconstruction (PR), balanced steady state free precession (SSFP) method for cardiac function evaluation against a two-dimensional (2D) multislice Fourier (Cartesian) transform (FT) SSFP method. The 3D PR SSFP sequence used projections in the x-y plane and partitions in z, providing 70-80 msec temporal resolution and 1.7 x 1.7 x 8-10 mm in a 24-heartbeat breath hold. A total of 10 volunteers were imaged with both methods, and the measurements of global cardiac function were compared. Mean signal-to-noise ratios (SNRs) for the blood and myocardium were 114 and 42 (2D) and 59 and 21 (3D). Bland-Altman analysis comparing the 2D and 3D ejection fraction (EF), left ventricular end diastolic volume (LVEDV) and end systolic volume (LVESV), and end diastolic myocardial mass (LVEDM) provided values of bias +/-2 SD of 0.6% +/- 7.7 % for LVEF, 5.9 mL +/- 20 mL for LVEDV, -2.8 mL +/- 12 mL for LVESV, and -0.61 g +/- 13 g for LVEDM. 3D interobserver variability was greater than 2D for LVEDM and LVESV. In a single breath hold, the 3D PR method provides comparable information to the standard 2D FT method, which employs 10-12 breath holds.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.