Abstract

Although intravascular ultrasound (IVUS) is an important tool in guiding complex coronary interventions, the resolution of existing commercial IVUS devices is considerably poorer than that of optical coherence tomography. Dual-frequency IVUS (DF IVUS), incorporating a second, higher frequency transducer, has been proposed as a possible method of overcoming this limitation. Although preliminary studies have shown that DF IVUS can produce complementary images, including large-scale morphology and high detail of superficial features, it has not yet been determined that this approach would be feasible in a more clinically relevant environment. The purpose of this study was to demonstrate the first in vivo use of a 30/80 MHz DF IVUS catheter in visualizing coronary vessels in a porcine model. In addition, two commercially available stents were studied in vitro and in vivo. Clear subjective improvement of visualization of superficial structures is demonstrated, and sufficient dynamic range is achieved to image through both the catheter sheath and blood in vivo.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.