Abstract

Background To evaluate the efficacy of a catheter system using a 3-Fr sheath with a steerable microcatheter through right upper limb artery access for superselective intra-arterial cisplatin infusion and concomitant radiotherapy (RADPLAT) to treat right maxillary sinus squamous cell carcinoma (MS-SCC). Material and methods We retrospectively studied 46 sessions in eight patients treated between November 2020 and February 2023 using the catheter system briefly described below. A 3-Fr sheath was inserted into the distal radial, conventional radial, or brachial arteries. A coaxial catheter system with a 2.9-Fr steerable microcatheter and a 1.9-Fr microcatheter was advanced into the brachiocephalic artery. The right common carotid artery was selected by bending the tip of the steerable microcatheter. Coil embolization and intra-arterial cisplatin infusion after selecting each external carotid artery branch were achieved using this catheter system. Results Cisplatin infusion and coil embolization were successful in all sessions. Arterial occlusion at the sheath insertion sites was found in 29.4% (5/17) of the distal radial arteries and 33.3% (3/9) of the conventional radial arteries. No other major complications were observed during the procedure. Conclusion Using a 3-Fr catheter system with a steerable microcatheter through right upper limb artery access is a feasible method for RADPLAT in treating right MS-SCC.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call