Abstract

To evaluate a 3-dimensional adaptive raw-data filter in reducing streak artifacts in low dose chest computed tomographic (CT) images. Fourteen adult patients who underwent low dose chest CT examination (parameters: 25 or 50 mAs, 120 kV) on 64-detector CTscanner were included in this study. We prepared 2 sets of contiguous 5-mm thick images by reconstruction with and without 3-dimensional adaptive raw-data filter (filter-processed and unprocessed images). Streak artifacts and visualization of peripheral vessels in both filter-processed and unprocessed images were evaluated using a 5-point scale. Upper, middle, and lower thorax were evaluated separately. The difference in artifact severity was statistically significant in upper and lower thorax (P = 0.002 and 0.03, respectively), whereas it was not significant in middle thorax (P = 0.13). The difference in the visibility of peripheral pulmonary vessels was not statistically significant in all anatomical regions. The 3-dimensional adaptive raw-data filter reduced streak artifacts in low dose chest CT in upper and lower thorax.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.