Abstract
Real-time 3-D intracardiac echocardiography (ICE) can enable faster imaging of surfaces orthogonal to the transducer, such as the pulmonary vein (PV) antra and cardiac valve annuli. However, the requirement for a 2-D grid of individually wired elements makes a traditional matrix array challenging to implement within an intravenous catheter. Helicoid array transducers are linear array transducers twisted about their long axis, allowing imaging of different elevation slices using sub-apertures. In this work, we examined the 3-D imaging characteristics of helicoid array transducers through simulations using Field II software and experimental measurements. We report results for varying transducer parameters, such as twist rate and sub-aperture size. We also discuss design considerations for these imaging parameters as they pertain to volumetric imaging of the heart.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.