Abstract

This paper proposes a calibration method for a freehand 3-D ultrasound system in medical robotic research. The calibration block with six wires was designed to set the fixed target points. The positions of the ultrasound probe and calibration block were measured using an optical tracker. The relationship between the position of the ultrasound probe and the pixel coordinates in the images was derived using a calibration process. The scaling matrix was also calculated. The experimental results showed that the proposed method could find solutions using a simple least square method from one or multiple ultrasound images.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.