Abstract

Intensity modulated radiation therapy (IMRT) delivery involves a complex series of beam angles and multileaf collimator (MLC) arrangements, requiring quality assurance to be performed to validate delivery before treatment. The purpose of this work is to investigate the effect of dose gradient on quality assurance (QA) passing rate. Many (n = 40) IMRT plans were delivered and measured using a 2D planar array of ion chambers; additionally, eleven plans were measured at several coronal planes. For each measurement, dose gradient was assessed using a number of metrics and passing rate assessed at both 3%/3 mm and 3%/2 mm criteria. The passing rates of the various IMRT plans were shown to be generally correlated to gradient, with an average distance correlation of 0.54 ± 0.04 for the lateral dose gradient. The passing rate for an individual plan was shown to vary with coronal slice, though the correlation to dose gradient was not predictable. Even though the passing rate was strongly related to dose gradient for many of the plans, the signs of the correlations were not always negative, as hypothesized. The coronal plane at which QA is performed affects passing rate, though dose gradient may not easily be used to predict slices at which passing rate is higher.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.