Abstract
Due to the features of modern radiotherapy techniques, such as Intensity Modulated Radiation Therapy (IMRT), Stereotactic Treatments with photons and proton therapy, where high spatial dose gradient are often present, detectors to be employed for two-dimensional dose verifications must satisfy narrow requirements. In particular, they have to exhibit high spatial resolution. For these applications, in the framework of the European Integrated project MAESTRO (LSHC-CT-2004-503564) and of the INFN experiment PRIMA, we designed a modular system based on a monolithic silicon segmented sensor. A single sensor has been coupled with readout electronics and tested with satisfactory results by using 6, 10 and 25MV X-rays from a LINAC at the University Hospital of Florence and 62MeV protons at INFN LNS Catania, following MAESTRO procedures. For photons, almost all the channels exhibit performances within project specifications (repeatability ≪0.5%, reproducibility ≪1%, deviation from linearity ≪1%, dose rate dependence ≪1%). For protons, the measured Spread Out Bragg Peak is in good agreement with the one measured with a single diode and the detector shows also a good linearity in the range 20–5000 cGy. The output factors are in agreement with those measured with ionization chamber, single diode or film, within experimental errors.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.