Introduction The INTRABEAM® system is a miniature accelerator producing low energy photons (50 keV maximum). The published dosimetric characterization of the INTRABEAM system for flat and surface applicators [1] , [2] was based on detectors (radiochromic films or ionization chambers) not allowing measuring the absorbed dose in the first millimeters of the irradiated medium, where the dose is actually prescribed. This study aims at measuring the dose deposit produced with INTRABEAM surface applicators in the first millimeters by determining gel sensitivity with Magnetic Resonance Imaging (MRI). Methods The irradiations at different dose levels were performed with the INTRABEAM® Carl Zeiss Surgical system (Oberkochen, Germany)]. The gel used in this study is a new “sensitis” material which is described by Stien et al. [3] and V. Dedieu et al. [4] . Gel irradiation in tin containers was carried out for twelve dose levels between 2 Gy and 50 Gy at the gel surface with a 4 cm surface applicator. In order to compare gel sensitivity at low energy with high energy, gels in tins containers were irradiated by a 18 MV photon beam produced by a Varian Clinac 2100 CD. T2 weighted multi echo MRI sequences were performed with 16 echo times and TR = 2000 ms. The response in T2 was determined on Matlab (Version 7.11, MathWorks, Natrick, MA, US). Results The T2 signal versus echo times can be fitted with a mono-exponential function with 95% of confidence. The calibration curve determined from experiments with tins at low energy is a linear function (R2 = 0.964) with a sensitivity of 0.0422 s−1 Gy−1. It is of the same order of magnitude as the gel sensitivity found from experiments at high energy (0.0391 s−1 Gy−1 with R2 = 0.924). The calibration curve at low energy was used to obtain Isodose maps from images (Fig. 1). The PDD (Percent Depth Dose) determined in gel is similar to the one measured with an ionization chamber. Conclusions The dosimetric response of ferrous gels in a low energy beam produced by a mini -accelerator was determined with MRI measurements. It allowed measuring the dose distribution in particular in the first millimeters for an irradiation with the INTRABEAM® miniature accelerator equipped with a surface applicator. These results must be confirmed for the other diameters of surface applicators and for flat applicators as well.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.