AbstractThe crossed‐loop resonator (CLR) uses two orthogonal lumped‐element resonators [one to excite the spins and one to detect the electron paramagnetic resonance (EPR)] to isolate the signal from the microwave source. It eliminates the need for a circulator. The high isolation provided by the CLR reduces the energy stored in the resonator that detects the signal, thereby reducing the intensity of the resonator ring down after the pulse, which decreases the instrument dead time. Overcoupling and synchronous switching of the Q's of the two resonators between high and low states were used to further reduce dead time and maximize the EPR signal following a pulse. Each section of a 250 MHz resonator that accommodates 1 inch diameter samples had a critically coupled Q of 950 when Q‐switching was not installed. With Q‐switching, free induction decays and electron spin echoes for 0.2 mM aqueous solutions of triarylmethyl radicals were obtained with dead times of a few hundred nanoseconds. Typically 50–60 dB isolation was achieved with various samples. © 2002 Wiley Periodicals, Inc. Concepts in Magnetic Resonance (Magn Reson Engineering) 15: 37–46, 2002.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.