Abstract

Abstract This manuscript presents characteristic dosimetry measurements for the 25 MV X-ray beam from the Varian Associates Clinac ® 35 linear accelerator. The quality of the beam is expressed by its attenuation as a narrow beam in lead. Percentage depth dose tables are given as a function of field size for square fields. Tissue-phantom ratios are tabulated for square field sizes. A study of the variation of the dose in the buildup region demonstrates that surface dose and the depth of maximum dose is a strong function of the field size. Other beam parameters measured and discussed include field flatness and symmetry, penumbra, and applicability of inverse square law.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.