Abstract

In the present work, electrochemical impedance immunosensor is developed for detection of myoglobin. The procedure involves the immobilization of myoglobin antibodies on the interdigitated gold electrodes (IDE) using alkanethiol self-assembled monolayer (SAM) for binding myoglobin antigens available in the blood serum/aqueous solution. Then the sensor system is characterised using alternating current (AC) electrochemical impedance spectroscopy (EIS), in which the change in impedance spectra is observed across the frequency range of 0.1 Hz to 10 MHz. This work mainly focuses on understanding and improving the impedance immunosensor for better sensitivity readings (~1 ng/mL) by adjusting the interdigitated electrode configuration, incubation times, reagent sample volumes and immobilisation protocols. Myoglobin is one of the premature indentifying cardiac protein markers to monitor minor heart attacks. Hence, the developed immunosensor has potential of using as a point-of-care diagnostic device. The protocol developed in this work can be useful for detecting other cardiac markers like Troponin and CK-MB by using respective antibodies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.