Abstract
Tube to tube volume difference presents a challenge in obtaining correct external standard quantitative NMR (esqNMR) results. Deuterium (2 H) NMR is easily observable, intrinsically quantitative, present in all samples, free of interfering signals, and insensitive to probe tune/match and sample saltiness. These properties make 2 H peak integral an ideal parameter in esqNMR for correcting volume differences between the reference standard and analyte. We demonstrate a novel and practical technique abbreviated as "2 H SOLCOR" (2 H SOLvent CORrected), where the 2 H peak integral from the solvent is used as a universal internal standard to correct volume variations in NMR tubes, thereby improving accuracy and precision of esqNMR method. Herein, this simple yet effective technique is described, and practical considerations for successful implementation are presented. 2 H SOLCOR can be applied anywhere esqNMR is used, including where precious samples need to be accurately quantified for qualification as an authentic analytical standard.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.