2,6-Pyridinedicarboxylic acid (2,6-PDC) was investigated as an eluent for ion chromatographic (IC) separation of inorganic anions, Mg2+ and Ca2+ followed by UV and conductivity detection. Other eluents containing trimesic, pyromellitic and commonly used gluconate/borate were also studied. Eluents containing 2, 6-PDC gave reasonable resolution and detection sensitivity for all analytes of interest. The effects of eluent pH and concentration of 2,6-PDC on the retention and sensitivity were investigated systematically. The separation and detection of HCO3−, H2PO4−, Cl−, NO2−, NO3−, SO4−, Mg2+ and Ca2+ within 11 minutes were performed using 1 mM 2,6-PDC at pH 6.0. Detection limits were in the range 5–15 μM for UV detection and 1.5–4 μM for conductivity detection. The proposed method offers a quick, sensitive and simple means for the determination of common anions, Mg2+ and Ca2+ particulaly in water samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.