Abstract

We report on an arrangement that measures angle resolved scattering (ARS), total scatter (TS), transmittance (T), and reflectance (R) at 157 nm and 193 nm. The ARS set-up is based on a high precision double goniometer arrangement, which can be inserted into the measurement chamber without removing the TS set-up. The TS set-up for detection of forward scatter and backscatter with extremely low background scatter levels of 1 ppm consists of an excimer laser, a Coblentz sphere with detection system, and a beam preparation path. The sphere and preparation path are housed in vacuum chambers allowing operation in vacuum or purge gas. The measurement options altogether constitute a multifunctional system: VULSTAR (<b>VUV</b> <b>L</b>ight <b>S</b>catter, <b>T</b>ransmittance, and <b>R</b>eflectance). We present total scatter measurements on deep ultraviolet (DUV) and vacuum ultraviolet (VUV) substrates and optical components with antireflective (AR) and highly reflective (HR) coatings, angle resolved measurements on optical components and R and T measurements on substrates for VUV optical coatings.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.