Abstract

Introduction The amorphous silicon (a-Si) electronic portal imaging devices (EPIDs) mounted on linacs have been extensively studied and have shown dose-response characteristics suitable for dose verification. Their use for dosimetric purposes is already in clinical routine. However, the response of EPID in continuous acquisition mode was not so studied in literature, and in less extent when considering it for dosimetric applications in stereotactic body radiotherapy (SBRT) conditions (i.e. small fields, and high dose rates using flattening filter free (FFF) beams). The purpose of this work was therefore to characterize the EPID response in terms of relevant dosimetric properties, mainly for the potential use of continuous mode for further SBRT dosimetry with dynamic techniques. Methods The EPID used in this study was the aSi-1000 model installed on a Varian TrueBeamSTx. Its response was characterized in FFF photon beams with nominal energies of 6 and 10 MV and maximum available dose rates (1400 and 2400 MU/min, for 6 and 10 MV respectively). Different sets of images were acquired in both integrated and continuous modes, with EPID at 150 cm from the source, to study dose linearity, repeatability and reproducibility of response, field-size dependence, ghosting effect and response to energy spectrum variations. Before data acquisition, dark field and flood filed images were acquired for offset and gain corrections. As fluctuations in dose rate can potentially influence the response of EPID when using dynamic techniques with continuous acquisition mode, the linearity of EPID response was further evaluated by irradiating a 10 × 10 cm2 field in 20° and 90° arcs focusing on low MU range (3–100 MU), and was also compared to static irradiations. Results Conclusions The characterization of the aSi-1000 EPID response to relevant dosimetric properties was investigated, showing the feasibility of its use for dosimetry with FFF beams and high dose rates. Results for continuous mode, which differed from dosimetric non-linearity reported in previous published studies [ [1] McCurdy B.M.C. Greer P.B. Dosimetric properties of an amorphous-silicon EPID used in continuous acquisition mode for application to dynamic and arc IMRT. Med Phys. 2009; 36: 3028-3039 Crossref PubMed Scopus (73) Google Scholar ], also encourage the use of EPID for SBRT with dynamic techniques in future works for 3D dosimetry applications (e.g. quality assurance including in vivo dosimetry). The amorphous silicon (a-Si) electronic portal imaging devices (EPIDs) mounted on linacs have been extensively studied and have shown dose-response characteristics suitable for dose verification. Their use for dosimetric purposes is already in clinical routine. However, the response of EPID in continuous acquisition mode was not so studied in literature, and in less extent when considering it for dosimetric applications in stereotactic body radiotherapy (SBRT) conditions (i.e. small fields, and high dose rates using flattening filter free (FFF) beams). The purpose of this work was therefore to characterize the EPID response in terms of relevant dosimetric properties, mainly for the potential use of continuous mode for further SBRT dosimetry with dynamic techniques. The EPID used in this study was the aSi-1000 model installed on a Varian TrueBeamSTx. Its response was characterized in FFF photon beams with nominal energies of 6 and 10 MV and maximum available dose rates (1400 and 2400 MU/min, for 6 and 10 MV respectively). Different sets of images were acquired in both integrated and continuous modes, with EPID at 150 cm from the source, to study dose linearity, repeatability and reproducibility of response, field-size dependence, ghosting effect and response to energy spectrum variations. Before data acquisition, dark field and flood filed images were acquired for offset and gain corrections. As fluctuations in dose rate can potentially influence the response of EPID when using dynamic techniques with continuous acquisition mode, the linearity of EPID response was further evaluated by irradiating a 10 × 10 cm2 field in 20° and 90° arcs focusing on low MU range (3–100 MU), and was also compared to static irradiations. The characterization of the aSi-1000 EPID response to relevant dosimetric properties was investigated, showing the feasibility of its use for dosimetry with FFF beams and high dose rates. Results for continuous mode, which differed from dosimetric non-linearity reported in previous published studies [ [1] McCurdy B.M.C. Greer P.B. Dosimetric properties of an amorphous-silicon EPID used in continuous acquisition mode for application to dynamic and arc IMRT. Med Phys. 2009; 36: 3028-3039 Crossref PubMed Scopus (73) Google Scholar ], also encourage the use of EPID for SBRT with dynamic techniques in future works for 3D dosimetry applications (e.g. quality assurance including in vivo dosimetry).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.