Abstract

The NOSAMS facility at Woods Hole Oceanographic Institution has started to develop and apply techniques for measuring very small samples on a standard Tandetron accelerator mass spectrometry (AMS) system with high-current hemispherical Cs sputter ion sources. Over the past year, results on samples ranging from 7 to 160 μg C showed both the feasibility of such analyses and the present limitations on reducing the size of solid carbon samples. One of the main factors affecting the AMS results is the dependence of a number of the beam optics parameters on the extracted ion beam current. The extracted currents range from 0.5 to 10 μA of 12C− for the sample sizes given above. We here discuss the setup of the AMS system and methods for reliable small-sample measurements and give the AMS-related limits to sample size and the measurement uncertainties.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.