Abstract

Study Objective To evaluate the accuracy of bladder scanner measurements in assessing post-void residual (PVR) volumes in obese women [body mass index (BMI)≥30 kg/m2] as compared to non-obese women (BMI Design Prospective. Setting Academic center. Patients or Participants Women undergoing multichannel urodynamic studies from June to September 2018. Interventions After uroflowmetry, PVR was measured with the BVI 3000. The largest volume of three attempts was recorded. Sterile straight catheterization was performed immediately after. The primary outcome was the difference between bladder scanner PVR and catheterized PVR in obese and non-obese women. Measurements and Main Results 133/157 women (57 obese, 76 non-obese) were eligible for inclusion. Obese women were younger (54.9±11.5 vs 63.0±11.8 years, p Conclusion There is no significant difference in PVR measurements by bladder scan versus catheterization in obese versus non-obese women.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.