Abstract

Dosimetric measurements close to radioisotope sources, such as those used in brachytherapy, require high spatial resolution to avoid incorrect results in the steep dose gradient region. In this work the Fricke Xylenol Gel dosimeter was used to obtain the spatial dose distribution. The readings from a 137Cs source were performed using two methods, visible spectrophotometer and CCD camera images. Good agreement with the Sievert summation method was found for the transversal axis dose profile within uncertainties of 4% and 5%, for the spectrophotometer and CCD camera respectively. Our results show that the dosimeter is adequate for brachytherapy dosimetry and, owing to its relatively fast and easy preparation and reading, it is recommended for quality control in brachytherapy applications.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call