Abstract

Our goal was to report a 10-year follow up of linear penetration rates for HCLPE, and to determine whether a difference exists between penetrations measured on pelvis or hip anterior-posterior radiographs. We reviewed 48 total hip arthroplasties where a first-generation HCLPE liner was used. Femoral head penetration was measured on both AP pelvis and hip radiographs. Total wear and wear rate at 10 years were 1.26 mm and 0.122 mm/y, respectively. The rate decreased significantly after the first 2-3 years, plateauing at a wear rate of 0.05 mm/y for the last 5 years. The AP hip total wear and wear rate were 1.38 mm and 0.133 mm/y respectively, while rates were 1.13 mm and 0.109 mm/y respectively for the pelvis radiographs (P<.05). We found a significant difference in measurements of linear penetration when comparing AP pelvis vs. hip radiographs with lower rates recorded using an AP pelvis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.