Abstract

We report the first definitive PM and AM noise measurements at 100 GHz of indium phosphide (InP) amplifiers operating at 5 K, 77 K, and room temperature. Amplifier gain ranged from +7 to +30 dB, depending on input RF power levels and operating bias current and gate voltages. The measurement system, calibration procedure, and amplifier configuration are described along with strategies for reducing the measurement system noise floor in order to accurately make these measurements. We compute amplifier noise figure with an ideal oscillator signal applied and, based on the PM noise measurements, obtain NF=0.8 dB, or a noise temperature of 59 K. Measurement uncertainty is estimated at /spl plusmn/0.3 dB. Results show that the use of the amplifier with an ideal 100-GHz reference oscillator would set a lower limit on rms clock jitter of 44.2 fs in a 20-ps sampling interval if the power into the amplifier were -31.6 dBm. For comparison, clock jitter is 16 fs with a commercial room-temperature amplifier operating in saturation with an input power of -6.4 dBm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.