Abstract
This paper presents a novel 0.4 V $G_{m}\mathrm{C}$ integrator design approach and fourth order bandpass filter design for EMG biomedical applications with high DC gain and and low power consumption. The filter and integrator are implemented with a new architecture in 180nm CMOS process. By using negative resistance in proposed integrator the response can be controllable. The negative resistance in this paper is implemented with Operational Trans Conductance Amplifier(OTA) The response of proposed integrator is close to ideal integrator behavior.The OTA used in this work is operated with 0.4 V supply and bulk driven structure. The maximum power consumption of integrator is 45 nW and cut off frequency is 160 Hz. The fourth order bandpass filter is consumes 117nW power with bandwidth 260 Hz. The dynamic range of the filter is 63dB.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.