Abstract

PPG signal measured by pulse oximeter can measure pulse and the oxygen saturation of arterial blood. But the PPG signal is distorted by finger movement or other movement in the body. To detect pulse from the PPG signal with motion artifact, we use band pass filter(BPF), Independent component analysis(ICA) and nonlinear autocorrelation(NAC). BPF is used to remove DC component and high frequency noise in the PPG signal with motion artifacts. ICA is used to separate pulse signal and motion artifact. However, pulse signal separated by ICA have no choice but to accompany signal distortion because pulse signal and motion artifact are not completely independent. So, we use nonlinear autocorrelation to emphasize the pure pulse signal from the distorted signal.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.