Abstract

In high static field magnetic resonance imaging(MRI) systems, fields of 7 T and 9.4 T, the impressed RF field shows larger inhomogeneity than in clinical MRI systems with B0 fields of 1.5 T and 3.0 T. In multi-channel RF coils, the magnitude and phase of the input to each coil element can be controlled independently to reduce the non-uniformity of the impressed RF field. The convex optimization technique has been used to obtain the optimum excitation parameters with iterative solutions for homogeneity in a selected ROI(Region of Interest). To demonstrate the technique, the multichannel transmission line coil was modeled together with a human head phantom at 400 MHz for the 9.4 T MRI system and fields are obtained. In this paper, all the optimized in each isolated ROIs are combined to achieve significantly improved homogeneity over the entire field of view. The simulation results for 9.4 T MRI systems are discussed in detail.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.