Abstract

A high-performance liquid chromatographic method was validated for quantitation of nadolol in human plasma. Nadolol and internal standard, pindolol, were extracted with tert-butyl methyl ether after addition of 10 M sodium hydroxide solution. The analytes were separated on a reverse phased C18 column using a mobile phase consisting of 0.05 M ammonium phosphate monobasic buffer, acetonitrile and methanol (81: 17:2 v/v/v) and detected using a fluorescence detector (excitation wavelength 230 nm, emission wavelength 294 nm). The method was specific and sensitive enough to detect as low as 3 ng/mL of nadolol in human plasma. Linear calibration range was 3-150 ng/mL with correlation coefficient greater than 0.999. The overall accuracy was in the range of 96.8 to 103% and precision C.V.(%) 7.30 to 12.2%. The recovery was approximately 100% and stability was confirmed during storage and sample preparation. The present HPLC method was successfully applied to study bioavailability after oral administration of 80 mg of nadolol in healthy Korean subjects. The mean was and of was reached at . The mean of nadolol was .

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.