Abstract

EO GRD(Ground Resolved Distance) resolution of airborne EO/IR(Electro-Optical/Infrared) sensor is a critical factor in test and evaluation for EO sensor performance. We propose the laboratory measurement set-up for EO GRD by constructing optical collimator which includes integrated sphere, blackbody, equivalent 3-bar target and 6 DOF motion simulator. GRD is measured in the photographic imagery of bar targets by 3 different distances for 3 EO/IR sensors and the measured results were analyzed statistically. We found that at least 7 sheets of imagery are needed in order to obtain meaningful EO GRD. The result of statistical analysis shows that the distribution of the measured GRD is nearly symmetric about the average GRD, and the better imagery ratio above the average GRD is about 40~70%. Also from the best GRD analysis, it is estimated that the design goal for EO GRD should be 30% superior to the required GRD.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.