Abstract

Immunoassays are used for detecting protein targets for various applications. Here, a modification of immunoassays to allow a purely electrical detection of the target protein concentration is shown. The modification comprises a β-D-glucosidase as reporter enzyme and a cyanogenic glycoside as substrate. The enzymatic reaction produces cyanide in small quantities. For electrical detection of the cyanide, a novel sensor is developed, based on a gold micro wire. The cyanide dissolves the gold wire and changes the electrical resistance of the wire. Monitoring the resistance change allows a quantitative measurement of the target human C-reactive protein (an inflammatory marker) in blood plasma in the physiological relevant concentration range.