Abstract

In this study, the effect of humidity change on DNPH cartridge sampling efficiency for carbonyl compounds (CCs) has been investigated. For this analysis, gaseous standard of 6 different CCs (formaldehyde (FA), acetaldehyde (AA), propionaldehyde (PA), butyraldehyde (BA), isovaleraldehyde (IA), and valeraldehyde (VA)) was calibrated after derivatization with three types of DNPH cartridge products. Their calibration results derived at RH values between 0 and 80% were then compared against liquid phase standards. If the results of our analysis are compared by the RH values between 20 and 80%, the effect of RH can be distinguished between light and heavy CCs. For lighter CCs (like FA and AA), there was no significant change. However, for the ones heavier than PA, there was fairly noticeable increase in relative recovery ratio in RH value between 20 and 80%. Such patterns are seen consistently from all three DNPH products tested for comparison. The results of our analysis suggest that proper correction for RH change may be needed for heavier CCs by the cartridge method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.