Abstract
The purpose of this study is to determine calculation accuracies of inhomogeneity correction using the absorption equivalent thickness (AET) method and the pencil beam method for electron dose distribution computations. l) In the AET method. AET factors are used to correct doses for tissue heterogeneties. In our experiment. they were well fitted to the values in the following way, AET=1.2 〓 (for lung tissue) AET=p〓 (for bone or soft tissue) where p 〓 Is the electron density of the tissue. The calculation error of the AET method in a lung tissue equivalent medium, cork, was more than 15% for electron energy of less than 10 MeV and/or small fields, since this method is not capable of taking into account the changes of electrons scattered from the other parts. 2) In the pencil beam method, computating results agreed well with the experimental ones within a maximum error of 10% for every energy range and every field.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.