In this 9th report, the influence of intensifying power on X-ray photographic contrast was studied ; high intensifyingscreen "Kyokko" HS and low intensifying screen "Kyokko" FS were used in the comparative study and examined by the tangent of the Al absorption wave within the tube voltage between 60kV and 120kV ; to verify the result, they were also examined in chest radiography. The following result wasproved ; degree of contrast under various intensifying powers depends not on the intensifying screen itself buton the condition of radiography in which the screen is used.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.