Abstract

The paper discusses the results of the calibration of two corona neutron counters used to measure the total neutron yield from the plasma of the Globus-M2 tokamak. The calibration was carried out in the experimental hall of the Globus-M2 facility using an AmBe source. During the calibration, the source moved uniformly around the central solenoid in the equatorial plane of the vacuum chamber, and one of the detectors was gradually moved away from the tokamak along a line with a constant toroidal angle. The values of the calibration coefficient obtained depending on the distance of the detector from the tokamak axis are presented. The calibration technique made it possible to separate in the detector signal the contributions from the direct neutron flux emitted by the plasma and from the flux of neutrons scattered on the elements of the experimental hall.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call