Abstract

Weighted residuals and the reduced chi(2) (chi(R) (2)) value are investigated with regard to their relevance for assessing optical property estimates using the diffusion equation for time-resolved measurements in turbid media. It is shown and explained, for all photon counting experiments including lifetime estimation, why chi(R) (2) increases linearly with the number of photons when there is a model bias. Only when a sufficient number of photons has been acquired, chi(R) (2) is a pertinent value for assessing the accuracy of mu(a) and mu(s)' estimates. It was concluded that chi(R) (2) is of particular interest for cases of small interfiber separation, low-level scattering, strong absorption and incorrect measurement of instrument response function. It was also found that chi(R) (2) is less pertinent for judging mu(a) in case of air boundary effects.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.