Abstract
In order to clarify the mechanism of high frequency titration, a bridge suitable for high frequency wave was constructed and was so devised that the dielectric constant of a solution possessing conductivity could also be measured. This made it possible not only to measure dielectric constants and loss of dielectrics with ease and rapidity but also to carry out titration with higher accuracy and without any limit to the concentration. It has also effected the combination of the so-called high frequency titration and conductivity titration. Cells insulated by glass was constructed, equation to calculate its cell constant was obtained, and actual measurement was carried out.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.