In this study, we performed computational fluid dynamic simulations to explore how the detailed design of drug-eluting stents affects both the flow field in the vicinity of the stent as well as the concentration of the eluted drug at the endothelial cell surface. Simulations were performed on three idealized stent geometries we developed and on geometries approximating three commercial stents,: Medtronic's Aurora stent, Cordis's BX Velocity stent, and Boston Scientific's Wallstent. An important contribution of the present study is the introduction of the stent effectiveness index (EI), which provides a quantitative assessment of stent performance and an objective basis for comparing the performance of different stents. Among the three commercial stents studied, our simulations have revealed that the BX Velocity stent is associated with the lowest in-stent EI values for the range of flow Reynolds numbers studied (). In addition to commercial stent designs, we investigated the EI in three idealized stents and determined that a spiral stent provides excellent performance (low EI) under all flow conditions investigated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.