Abstract

To improve the accuracy of Ritchey-Common test for a flat mirror, a new method to use the relationship between the system pupil coordinate and the test mirror coordinate to interpolate and fit the test mirror surface was proposed. On the basis of the least square method, the system defocus error and surface error in two test angles were detached to gain a more actual flat surface error. The simulation analysis shows that the test error can be controlled under /100(=632.8 nm). A flat mirror with a diameter of 40 mm was tested. In the test process, the test flat mirror was rotated with different angels and the wavefront was reconstructed through the relationship between coordinate mapping and amplitude conversion. The obtained results show that the RMS of the test mirror surface is 0.018 6 after detaching the defocus error of the system. As compared with the test RMS of 0.021 from an interferometer, the residual error is 0.002 4 . Experiment results indicate that this error detaching method is valid and accurate in the Ritchey-Common test for flat mirrors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.