Abstract

Standard uroflowmetry measures the urine weight using single load cell to evaluate the urinary flow rate. Impact noise should be introduced due to gravity when the urine stream falls down into the container upon the load cell. The present study placed three load cells on the three vertices of a regular triangle and the three signals were ensemble averaged to enhance the signal-to-noise ratio(SNR) regardless of how the urination was made. Simulated urination experiment was performed with three different urine collection methods. In all three methods, SNR of the averaged signal was much higher than each load cell signals. With no urine collection device, the present signal averaging technique resulted in SNR values higher by 10~15 dB than when dual funnels or upper funnel were used to guide the urine stream. Therefore, it was demonstrated that the three point measurement followed by with ensemble averaging could enable accurate uroflowmetric test without any specially made urine collection devices.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.