Abstract

We proposed and optimized a simple method of temporal subtraction image between successive bone single photon emission computed tomography (SPECT) images for supporting interpretation of temporal changes, and we evaluated its clinical utility. This method consisted of image registration, count normalization, and image subtraction. For image registration, we used a BEAT-Tl software. For count normalization, a pixel value of the normal accumulation part in a SPECT image was used as a reference region. We evaluated accuracy of image registration and optimized the normalization procedure. The accuracy of image registration ranged within 1 pixel in all directions (x, y, x-axis, and rotation). As the reference region, the second lumbar vertebra showed the best results in terms of the normalization procedure. Our method simply allowed the production of a temporal subtraction image. Because the software used in this method can be used free, this method would be available in every institution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.