Abstract
The gross alpha analysis of nasal smear samples for the radiation emergency and the additional follow-up steps were established. Cotton swab sticks using in local hospitals for nasal smear in Korea were used for the verification. The measurement results of standard samples spiked with certified reference source were well agreed within compared with reference values. The clearance ratio of smear samples conducted with wet smear condition showed higher value. To eliminate the quenching effect of liquid scintillation samples, dry of smear samples should be followed up before counting samples. Based on the measurement results, medical decision levels and internal dose assessment were established for the victims in the beginning of radiation emergency.
Paper version not known (Free)
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.