The master-slave neurosurgical robotic system MM-3 was developed to assist with difficult tasks in neurosurgery. In this system, the hand motion of the operator is scaled down using the motion-scaling ratio. In this paper, we propose the modulation of the motion-scaling ratio based on microscope magnification. First, an experiment for a simple task was conducted by engineering students, and the result showed that the proposed method reduced both the task completion time and path length of the master manipulators in a coarse motion without compromising usability. However, the result of suturing experiment by neurosurgeons did not show the effectiveness of the proposed method. The method needs to be evaluated in a task that can be performed without errors. Customization of the motion-scaling modulation method according to individual doctor's preference needs to be investigated in the future.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: The Proceedings of JSME annual Conference on Robotics and Mechatronics (Robomec)
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.