The <TEX>$Comet{\acute{e}}$</TEX> Consultatif de <TEX>$Thermom{\acute{e}}trie$</TEX> (CCT) under the Comete International des Poids et Measures (CIPM) has decided to perform the Key Comparison (KC) for triple point of water cells used as a reference fixed point of thermometry at the 21st meeting held at November 2001, and the Bureau International des Poids et Measures (BIPM) has been nominated as a KC coordinator. According to the KC protocol prepared by BIPM, KRISS performed the KC experiments and evaluate a uncertainty. The temperature difference between two reference cells for the Korea Research Institute of Standards and Science (KRISS) and a test cell for the transfer standard, which is moved to BIPM was 0.024 mK and the combined standard uncertainty evaluated 0.055 mK.