Abstract

The <TEX>$Comet{\acute{e}}$</TEX> Consultatif de <TEX>$Thermom{\acute{e}}trie$</TEX> (CCT) under the Comete International des Poids et Measures (CIPM) has decided to perform the Key Comparison (KC) for triple point of water cells used as a reference fixed point of thermometry at the 21st meeting held at November 2001, and the Bureau International des Poids et Measures (BIPM) has been nominated as a KC coordinator. According to the KC protocol prepared by BIPM, KRISS performed the KC experiments and evaluate a uncertainty. The temperature difference between two reference cells for the Korea Research Institute of Standards and Science (KRISS) and a test cell for the transfer standard, which is moved to BIPM was 0.024 mK and the combined standard uncertainty evaluated 0.055 mK.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call