Abstract

Crocidolite quality control (QC) sampling created by the wet generation method has been validated by validation tests such as the accuracy, precision, and storage tests. For this study we designed and developed a manufacturing apparatus and standard operating procedure for making these QC samples. The most important step in the procedure of making QC samples was the stage eliminating static electricity in asbestos fibers. This static electricity hampers the fibers clog functioning. In accuracy and precision tests by phase contrast microscopy analysis, the difference between the reference values and the studied values was at maximum 17.8%. This satisfies the AIHA proficiency analytical test criteria for asbestos. We could confirm the nearly even distribution of crocidolite fibers on the membrane filter. Also, there was no loss of fibers in the storage test after the one month.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.