Abstract

To compare the stationary dental X-ray generator and the portable dental X-ray generator and to understand spatial radiation dose depended on locations by measuring spatial radiation dose of the portable dental X-ray generator. The researchers used an Ionization chamber to measure spatial radiation dose which was generated while applying X-ray radiation to real bone skull phantom with both portable and stationary dental X-ray generator. There were 4 measurement locations which were immediate anterior, right, left and posterior. Distance of measurement was 50 cm in every location and the recorded result is an average of two applications of X-ray radiation to the maxillary molar area under the condition of 70 kVp, 3 mA, 0.1 sec. Average spatial radiation dose of portable X-ray generator was , much higher than that of stationary X-ray generator which was (p to (p and immediate right was the lowest of . Immediate left and posterior, however, resulted in similar radiation dose which were , (p