Abstract

Abstract Digital X-ray imaging devices using a TFT based flat panel array has been used in medical field. But, recently, lots of research on the photon counting sensor has been reported. In this study, we evaluated the physical properties of the photoconductor by suggesting the standard and testing method for quantitative performance evaluation of photon counting x-ray imaging sensor. First, we measured the leakage current and the sensitivity of photon counting x-ray imaging sensor and we evaluated the characteristic of rising time for determining the signal shaping time. In addition, the set-up study was conducted on the basis of the IEC 62220-1-2 recommendations to define the number of incident photons per unit area. And the efficiency of the charge collection was also evaluated. As a result, the characteristic was measured as 200 pA/mm 2 of the leakage current, 7 μC/cm 2 R of the X-ray sensitivity, and 0.765 μs of the rising time. Key Words : photon counting, photoconductor, x-ray sensor, charge collection efficiency

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.