Abstract

The focal spot of electron beam depending on the anode angle in the structure and major parts of the X-ray tube was investigated by the OPERA-3D/SCALAR simulation program. The simulation worked on four spaces with with two spaces, including anode and cathode of X-ray tube, by applying the finite element method analysis. The analytical model and dimension for the emission orbit of thermal electrons made from one filament of the focused X-ray cathode is affected to the penumbra of detector for the X-ray depending on any real focal spot size. The model shape of focusing cap and focusing tube with an anode target angle and a cathode filament is analyzed by the current density distribution of thermal electrons. The focusing width of thermal electrons for the X-ray tube depended on the anode angle (θ). The focusing value of electron beams at a region of anode angle having 10 ~17 maintained to below value of 70 μm. The minimum focal size of the electron beam was 40 μm at an anode angle of 15 . The focused X-ray tube of many variables depended on the thermionic emission of hot electrons from the target trajectory. The focusing tube will contribute to