Abstract
Neutron dose equivalent rates are measured in the irradiation room and the maze of a medical electron accelerator facility. An m-counter used for measurement is associated with a gate circuit to reject piled up signals caused by pulsed X-rays, so that it is possible to measure the neutrons without obstruction by the piled up signals even in the irradiation room. Counting loss of neutrons is mathematically corrected, and the correction factor comes nearly equal to 1.5 at dose points close to the entrance to the irradiation room when 14 MV X-rays are generated. The measurement in the maze agrees with the calculation reported in the preceding report in a factor less than 1.5 except for dose points close to the entrance to the irradiation room.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.