Abstract

In order to measure the pressure distribution on supporting structures under complete denture, we used impedance grid pressure sensor and examined its accuracy and availability. The results were as follows: 1. The sensitivity of this sensor was high when the distance between electrodes was short. 2. This sensor showed large change in output voltage and high sensitivity in the case of soft elastic materials. 3. This sensor showed the high linearlity between output voltage and applied load. 4. This sensor showed hysteresis loop under cycle load. 5. Base line of this sensor that has the urethane rubber as elastic material was changed when the load was applied intermittently. We measured the pressure distribution over supporting bones under upper complete denture by using simulator of maxillary edentulous with this impedance grid pressure sensor. We could observe the pressure distribution and pressure region in detail and visually when the load was applied to the both occlusal tables evenly.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.