Recently, As people's interest in the health of teeth is increased in the medical field changed into aging society, the number of times for the radiological diagnosis is increased. It can be said that the radiation exposure dose of Korean population is increased. It is also growing concern about radiation exposure. Therefore, the basic data for the dental panoramic X-ray system, its investigation and measuring the radiation dose is needed. In this study, we used ALOKA PDM-117 dosimeter and estimated a two-dimensional dose distribution of the dental panoramic X-ray system (VATEC Pax-400). Dose evaluation about the distribution is confirmed from the point of radiation exposure of a patient. Dose distribution of the dental panoramic X-ray system irradiated chin and the facial region to high dose as well as the parts of teeth. It was founded that the eye lens which are sensitive to radiation are exposed to unnecessary radiation, considering the effect of scattered radiation. The results of this study will be used more accurate dose assessment in a variety of object size and location of measuring dose. Keyword : X-ray, Exposure, Panoramic System, Eye